Application of a simple phantom in assessing the effects of dose reduction on image quality in chest radiography

Biomedical Physics and Bioengineering, University of Aberdeen, Foresterhill, AB25 2ZD, Aberdeen, UK

Received 21 March 2009; received in revised form 18 September 2009; accepted 26 September 2009.

Abstract 

Purpose

Firstly, to evaluate a commercial chest phantom incorporating a quasi anthropomorphic insert by comparing exposure measurements on the phantom with those of actual patients and, secondly, to assess the value of the phantom for image quality and dose optimisation.

Methods

In the first part of the study entrance surface doses (ESD), Beam transmission (BT), and optical density (OD) were obtained for 77 chest radiography patients and compared with measurements made from exposures of the phantom using the respective patient exposure factors from chest examination. Differences were assessed with a student t-test, while the Pearson's linear correlation coefficient was used to test for any linear relationship. The second part assessed the applicability of the phantom to image quality studies by investigating the effect, on the clarity and detectability of lung lesions made from gelatine, of reducing patient dose below current dose levels. Clarity of linear objects of different dimensions was also studied. Lesion detectability and clarity was assessed by four observers. The possibility of extending dose reduction below current dose levels (Dref) was assessed from comparison of doses that produced statistically significant differences in image quality from Dref.

Results

Results show that, with the exception of entrance doses and beam transmission through the diaphragm (P > 0.05), differences in OD and beam transmission between patients and phantom were statistically significant (P < 0.05). A Pearson's correlation test showed good positive correlation in ESD (r = 0.9) and beam transmission (r = 0.8) for all regions and a weaker positive correlation (r = 0.3) for OD in all areas. In both cases the correlation was statistically significant (P < 0.05). Extending dose reduction by between 41 and 65% below Dref produced significant changes in both clarity and detectability.

Conclusion

Within limits posed by the observed differences, the phantom can be applied to image quality studies in diagnostic radiology.

Keywords: Phantom, Chest radiography, Image quality, Clarity, Detectability, Dose reduction

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S1078-8174(09)00092-3

doi:10.1016/j.radi.2009.09.007

Radiography
Volume 16, Issue 2 , Pages 108-114, May 2010