Validation of an external ultrasound device for bladder volume measurements in prostate conformal radiotherapy

  • Evelyn O'Shea

      Affiliations

    • Department of Radiation Oncology, St. Luke's Hospital, Highfield Road, Rathgar, Dublin 6, Ireland
    • Corresponding Author InformationCorresponding author. Tel.: +353(0) 1 4065107; fax: +353(0) 1 4974886.
  • ,
  • John Armstrong

      Affiliations

    • Department of Radiation Oncology, St. Luke's Hospital, Highfield Road, Rathgar, Dublin 6, Ireland
  • ,
  • Tom O'Hara

      Affiliations

    • Infosight Training and Consultancy, Midlands Innovation and Research Centre, Athlone, Ireland
  • ,
  • Louise O'Neill

      Affiliations

    • Department of Physics, St. Luke's Hospital, Dublin, Ireland
  • ,
  • Pierre Thirion

      Affiliations

    • Department of Radiation Oncology, St. Luke's Hospital, Highfield Road, Rathgar, Dublin 6, Ireland

Received 19 September 2006; received in revised form 23 March 2007; accepted 7 June 2007.

Abstract 

Purpose

The objective of this study was to assess the feasibility and accuracy of an external ultrasound device – BladderScan™ 6100 instrument (BVI) – for the measurement of bladder volume for prostate conformal radiotherapy patients.

Materials and methods

The bladder volume delineated on the planning CT scan was compared with the volume measurements obtained from a non-invasive BVI, for 50 prostate conformal radiotherapy patients, using a paired t-test at 5% significance level. The BVI repeatability was calculated using the within-patient standard deviation of the BVI replicate measurements.

Results

This study demonstrates that bladder volume measurements obtained using the BVI are not significantly different from the volumes delineated on the planning CT scan, mean difference=−9.65ml, p=0.351. The magnitude of the difference between the CT and BVI measurements was found to be unrelated to the magnitude of the bladder volume. The repeatability of the BVI measurements was not influenced by the magnitude of the bladder volume.

Conclusion

The results confirm the validity of using the BVI to instantly, objectively and non-invasively measure a patient's bladder volume for prostate conformal radiotherapy. The validation of this instrument will permit further research to be conducted with regards to developing evidence-based bladder filling instructions for radiotherapy patients.

Keywords: Bladder volume, Prostate radiotherapy, Full bladder

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S1078-8174(07)00058-2

doi:10.1016/j.radi.2007.06.001

Radiography
Volume 14, Issue 3 , Pages 178-183, August 2008