Radiography
Volume 13, Issue 3 , Pages 214-220, August 2007

Dental radiography technique and equipment: How they influence the radiation dose received at the level of the thyroid gland

School of Health Sciences, University of Ulster, Shore Road, Newtownabbey, Belfast BT37 0QB, UK

Received 6 August 2005; accepted 3 March 2006.

Abstract 

Purpose

The aim of this study was to investigate the influence that collimator and technique choice had on the radiation dose detected at the thyroid gland position, during intra-oral examinations of the upper and lower teeth. Radiation dose reduction from a different perspective, other than the application of lead–rubber shielding, was addressed.

Methods

A study was performed at a regional dental school with the use of a phantom head/neck and a radiation dosemeter, to measure the radiation dose detected at the thyroid gland position. The radiation dose was assessed for two intra-oral techniques (paralleling and bisecting angle), and two collimators (rectangular and circular). The radiation dose was also assessed with and without the application of a thyroid shield. Standard descriptive statistics, followed by inferential statistics were applied to the data.

Results

There was a significant reduction in the radiation dose detected at the thyroid gland position, when employing the paralleling technique (66.7%) and rectangular collimator (45.5%). Other factors, for example the tooth/teeth under examination, were also found to influence the radiation dose detected.

Conclusion

Radiation dose reductions using the paralleling technique and rectangular collimator were outlined. The use of this low dose combination within dental practices remains limited, therefore, continued awareness and acceptance of radiation hazards need to be addressed.

Keywords: Oral radiology, Radiation dose reduction

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S1078-8174(06)00035-6

doi:10.1016/j.radi.2006.03.002

Radiography
Volume 13, Issue 3 , Pages 214-220, August 2007