A comparison of low contrast performance for amorphous Silicon/caesium iodide direct radiography with a computed radiography: A contrast detail phantom study

  • Mark McEntee

      Affiliations

    • Diagnostic Imaging, School of Medicine and Medical Science, UCD Dublin, Belfield, Dublin 4, Ireland
    • Corresponding Author InformationCorresponding author. Tel.: +353 1 7166356; fax: +353 1 7166547.
  • ,
  • Helena Frawley

      Affiliations

    • Beaumont Hospital, Dublin, Ireland
  • ,
  • Patrick C. Brennan

      Affiliations

    • Diagnostic Imaging, School of Medicine and Medical Science, UCD Dublin, Belfield, Dublin 4, Ireland

Received 1 September 2005; accepted 2 January 2006.

Abstract 

A range of digital image acquisition devices exists in diagnostic radiology. This study compares contrast performance of two such systems: an amorphous Silicon/caesium iodide (a-Si:CsI) based flat panel (DR) digital chest radiography system and a computed radiography (CR) system. Images of a contrast detail resolution phantom were acquired at a range of radiation doses. Three observers assessed all hardcopy images using a four-alternative forced choice observer perception technique. Contrast detail performance was calculated and low contrast performance quantified.

The DR system demonstrated significantly better low contrast performance and potential dose savings of up to 75% compared to the CR system. Threshold levels of contrast detail resolution were defined and levels of under- and over-exposure, compared to the threshold level, were highlighted. Both systems were noise limited at lower exposures and latitude limited at higher exposures. The results demonstrate that the DR system should perform better than the CR system under typical clinical conditions relevant to chest radiography particularly for the detection of low contrast details such as lung metastases or pneumothoraces.

Keywords: Computed radiography, Direct radiography, Contrast detail resolution, Dose efficiency

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S1078-8174(06)00005-8

doi:10.1016/j.radi.2006.01.004

Radiography
Volume 13, Issue 2 , Pages 89-94, May 2007